The sampler collects NO2 by molecular diffusion along an inert tube to an absorbent, in this case triethanolamine. The sampler, which is shown in the following diagram, is composed of a polypropylene housing with an opening of 20 mm diameter. To reduce wind disturbance a teflon membrane is attached, supported by a wire net. The collected NO2 is determined spectrophotometrically by the Saltzmann method.

The samplers are placed in a special shelter to protect them from rain.

Diagram

CLD Monitor

The diagram shows the linearity of the calibration function under laboratory conditions.

 

Specifications

Samplingrate

15.5  ml per minute at 20oC

Working range

5 – 240 µg/m3

Sampling time

8 to 48 hours

Detection limit

for a    8-hours exposure                 5µg/m3
for a 24 8-hours exposure               2 µg/m3

External influences:
   wind speed
influence of wind speed < 10% up to 10 m/sec
< 0.2 m/sec starvation
   temperature
no influence between 10 to 30oC
   humidity
no influence between 20 to 80%

Storage

before use: 6 months
after use:    6 months
Travelblanc recommended

Interferences

oxydants

Expanded uncertainty*

28.2 % at a level of 80 µg/m3

*according to GUM;
subject to change without notice
revised 3.1.2010

 

 