Benzene
 

ORSA5 diffusion tube comprises a glass sampling tube open at both ends and filled with activated charcoal. Each opening in the sampling tube is filled with a cellulose acetate diffusion barrier as the following figure shows:
Ambient air diffuses into the sampling tube in a controlled manner. BTX are absorbed on the activated charcoal and desorbed by carbon disulphide in the laboratory and analysed by gaschromatography.

Diagram

Comparison of ORSA diffusion tube with active measurement in North-Rhein Westfalia [1]

Specifications

Samplingrate

6.44 ml/min at 20oC

Working range

0.5 — 50 µg/m3

Sampling time

2 — 4 weeks

Detection limit

0.2 µg/m3 for a sampling time of 1 month

External influences:
   wind speed
influence of wind speed < 10% up to 4.5 m/sec
using protection shelters
   temperature
no influence between 10 to 30oC
   humidity
no influence between 20 to 80%
caution under tropical conditions

Storage

before use:      12 months
after exposure:  6 months

Cross sensitivity

Selective method

Expanded uncertainty*

23.0 % at concentration levels of 1 - 5 µg/m3

*according to GUM;
subject to change without notice
revised 3.1.2010

 

 
2007 Copyright by passam ag